Determination of Total Organic Carbon in process water

TOC analysis allows the quantification of Total Organic Carbon in galvanic process waters. Monitoring this parameter in water is important to ensure the high quality standards of the electroplating industry required by the market.

Method

Test type: End Point.

Testing time: 10 minutes.

Analysis sessions with multiple samples are possible up to a maximum of 16 tests for CDR GalvanLab® and 3 tests for CDR GalvanLab® Junior.

13%

Sample treatment

The sample can be used as is.

Principle of the test

Total Organic Carbon content is determined by its reaction with a chromogen that develops a coloration, the optical intensity of which is proportional to the concentration of Organic C in the analyzed sample.

Calibration curve

Calibration curve was obtained by correlating the CDR method with the reference method.

TOC Analysis in Process Water: Calibration Curve of the CDR GalvanLab® instrument with the reference method.

Reagent test Kits

Code *400140: 100 test
Code *400145: 10 test
The reagents have a shelf-life of 18 months

Measuring range

AnalysesMeasuring rangeResolutionRepeatability
Total Organic Carbon [5-100] mg/L5 - 100 mg/L1.000 mg/L5 mg/L

The chemical analysis instrument for electroplating baths

CDR GalvanLab®

  • Complete analysis panel, supplied already configured
  • Up to 16 determinations simultaneously
  • Possibility of carrying out analyses of the same sample
  • Integrated printer
  • Full connections (LAN - USB - Bluetooth barcode/QR code reader)

CDR GalvanLab® Jr

  • Partial analysis panel, customisable configuration
  • Up to 3 determinations simultaneously
  • Wireless connection to external printer
  • USB connections