Determination of Palladium concentration in Pd-Ni plating bath
The analysis allows the quantification of palladium concentration in Pd-Ni electroplating baths and recovery waters. Monitoring the amount of palladium in electroplating baths is essential to set the correct parameters (e.g. deposition time, deposition thickness) in the electroplating process ensuring its good performance and preventing aesthetic and/or functional defects in finished products.
Method
Test type: End Point.
Testing time: 4 minutes.
Analysis sessions with multiple samples are possible up to a maximum of 16 tests for CDR GalvanLab® and 3 tests for CDR GalvanLab® Junior.
Sample treatment
The sample can be used as is.
Principle of the test
Palladium content is determined by its reaction with a chromogen that develops a coloration, the optical intensity of which is proportional to the concentration of Pd in the analyzed sample.
Calibration curve
Calibration curve was obtained by correlating the CDR method with the reference method.
Reagent test Kits
Measuring range
| Analyses | Measuring range | Resolution | Repeatability |
|---|---|---|---|
| Palladium PdNi bath [3-8] g/L | 3 - 8 g/L | 0.010 g/L | 0.09 g/L |
| Palladium PdNi bath [0.3-4] g/L | 0.3 - 4 g/L | 0.010 g/L | 0.03 g/L |
The chemical analysis instruments for electroplating baths
CDR GalvanLab®
- Complete analysis panel, supplied already configured
- Up to 16 determinations simultaneously
- Possibility of carrying out analyses of the same sample
- Integrated printer
- Full connections (LAN - USB - Bluetooth barcode/QR code reader)
CDR GalvanLab® Jr
- Partial analysis panel, customisable configuration
- Up to 3 determinations simultaneously
- Wireless connection to external printer
- USB connections