Determination of Cyanide concentration in process water

The determination of total cyanides in electroplating process water is essential for monitoring plant efficiency and ensuring environmental safety. It is a key parameter for assessing the need for abatement treatments prior to discharge, in line with current regulations.

Method

Test type: End Point.

Testing time: 8 minutes.

Analysis sessions with multiple samples are possible up to a maximum of 16 tests for CDR GalvanLab® and 3 tests for CDR GalvanLab® Junior.

7%

Sample treatment

The sample can be used as is.

Principle of the test

The content of cyanides is determined, after dilution, by reaction with a chromogen. The test has a total duration of 10 minutes at the end of which the signal related to the colored complex formed is detected.

Calibration curve

Calibration curve was obtained by correlating the CDR method with the reference method.

Total Cyanides Analysis in Process Water: Calibration Curve of the CDR GalvanLab® instrument with the reference method.

Reagent test Kits

Code *400060: 100 test
Code *400065: 10 test
The reagents have a shelf-life of 12 months

Measuring range

AnalysesMeasuring rangeResolutionRepeatability
Cyanides [0.05-5] g/L0.05 - 5 g/L0.001 g/L0.043 g/L

The chemical analysis instrument for electroplating baths

CDR GalvanLab®

  • Complete analysis panel, supplied already configured
  • Up to 16 determinations simultaneously
  • Possibility of carrying out analyses of the same sample
  • Integrated printer
  • Full connections (LAN - USB - Bluetooth barcode/QR code reader)

CDR GalvanLab® Jr

  • Partial analysis panel, customisable configuration
  • Up to 3 determinations simultaneously
  • Wireless connection to external printer
  • USB connections